CA80TN Total Nitrogen Measurement Manual Overview

CA80TN is an analyzer used for online monitoring of total nitrogen concentration in water bodies, widely applied in wastewater treatment plants, surface water monitoring stations, and industrial wastewater discharge outlets. This manual provides complete guidance on installation, operation, calibration, and maintenance to ensure accurate and reliable measurement. You can click the link below to download the complete PDF document.

CA80TN Chinese Manual Download Download CA80TN Chinese Manual.pdf

Product Core Features

  • High-precision measurement: Uses UV digestion-colorimetric method, with a detection limit as low as 0.1 mg/L.
  • Automatic calibration and cleaning: Supports scheduled automatic calibration and cleaning, reducing manual maintenance.
  • Wide range options: Provides multiple range configurations such as 0-10 mg/L, 0-50 mg/L, 0-200 mg/L.
  • Data storage and output: Built-in large-capacity memory, supports 4-20mA, RS485 Modbus output.

Technical Parameters Overview

Parameter Item Specification
Measurement Principle UV digestion-colorimetric method
Measurement Range 0-10/50/200 mg/L (optional)
Accuracy ±5% FS
Repeatability ≤3%
Response Time ≤15 minutes
Output Signal 4-20mA, RS485 Modbus
Power Supply 220V AC ±10%, 50Hz
Protection Rating IP65 (transmitter)

Installation and Commissioning Steps

Pre-installation Preparation

  • Confirm that on-site power, signal cables, and sampling pipelines are in place.
  • Check the instrument appearance for damage and whether accessories are complete.
  • Read the safety warnings in the manual to ensure compliance with local electrical codes.

Installation Process

  1. Fix the analyzer on a bracket or wall to ensure it is vertical and stable.
  2. Connect the sampling tube to the water inlet and the drain pipe to the drain outlet, avoiding bending.
  3. Connect to 220V power supply and connect the 4-20mA or RS485 signal cable.
  4. Turn on the power and observe whether the display starts normally.

Commissioning and Calibration

  1. Enter the menu to set the range, output mode, and calibration cycle.
  2. Use standard solutions for zero and span calibration.
  3. Start the automatic cleaning program and confirm the pipeline is unobstructed.
  4. Record calibration data and save parameters.

Daily Maintenance and Troubleshooting

  • Regular reagent replacement: Based on usage frequency, usually replace every 1-3 months.
  • Clean the colorimetric cell: Clean at least once a month to prevent dirt from affecting measurement.
  • Check pipelines: Check weekly whether the sampling tube and drain tube are blocked or leaking.
  • Common faults: If readings are abnormal, first check whether the reagent has expired, whether the colorimetric cell is contaminated, and whether calibration is overdue.

Download the Complete Manual

For more detailed installation dimensions, wiring diagrams, menu operations, and fault codes, please download the complete version of the "CA80TN Chinese Manual".