Product Overview

MMP60 Material Moisture Analyzer is a precision instrument used for online measurement of material moisture in industrial sites. It adopts advanced sensing technology and can output moisture data in real time, suitable for grain, chemical, building materials, and many other industries. This Chinese manual aims to help users correctly install, operate, and maintain the MMP60 to ensure long-term stable operation.

Download MMP60 Chinese Manual.pdf

Technical Specifications

The following are the main technical indicators of the MMP60 for selection and acceptance reference.

ParameterSpecification
Measurement range0~100% RH (depending on material)
Accuracy±1% RH
Output signal4-20mA / RS485
Power supply24VDC
Protection classIP65
Operating temperature-20~60℃

Installation and Wiring

Installation Location

Select a highly representative measurement point, avoiding vibration, strong electromagnetic interference, and corrosive gases. The sensor should be fully inserted into the material to ensure good contact.

Electrical Wiring

  1. Confirm the power supply is 24VDC and the polarity is correct.
  2. Connect the 4-20mA output to a PLC or display instrument.
  3. Use shielded twisted pair for RS485 communication lines, connecting A/B correspondingly.
  4. After wiring, tighten the cable gland to ensure the protection class.

Operating Steps

Correct operation is key to ensuring measurement accuracy. Please follow the process below.

  1. After power-on, wait 30 seconds for warm-up.
  2. Enter the calibration menu via buttons or host computer.
  3. Use standard samples for zero and span calibration.
  4. Set the corresponding moisture range for the output.
  5. Return to measurement mode and record stable readings.

Maintenance and Troubleshooting

Daily Maintenance

  • Regularly clean the sensor surface to avoid material adhesion.
  • Calibrate accuracy every 6 months.
  • Check whether the cable seal is aging.

Common Faults

  • No output: Check power supply and wiring.
  • Reading drift: Recalibrate or clean the sensor.
  • Communication failure: Check address and baud rate settings.

Precautions

It is strictly prohibited to use non-explosion-proof MMP60 in explosive environments. Cut off the power before installation and maintenance. For the complete manual, please download the PDF file above.