VEGASOURCE 84 Chinese Manual Download

PDF download icon Click to download VEGASOURCE 84 Chinese Manual.pdf to get the complete operation and safety guide.

Product Overview

VEGASOURCE 84 is a radioactive source container used for industrial level measurement, usually used in conjunction with VEGAMET signal processing instruments. Its design is compact and its shielding is reliable, making it suitable for harsh working conditions such as high temperature, high pressure, and corrosion.

Technical Parameters

ParameterSpecification
Applicable source typeCs-137, Co-60, etc.
Shielding materialLead, tungsten alloy
Operating temperature-40°C ~ +80°C
Protection classIP66/IP67
Process connectionFlange or thread

Installation Steps

  1. Confirm that the installation location meets radiation safety requirements and set up warning signs.
  2. Use suitable lifting equipment to secure the container to the bracket or flange, ensuring the vertical or horizontal direction is correct.
  3. Connect the electrical wiring, paying attention to grounding the shielded wire.
  4. Check all fasteners to ensure there is no looseness.

Calibration and Commissioning

Zero Calibration

In the empty tank state, perform zero calibration through the VEGAMET instrument to eliminate the influence of background radiation.

Span Calibration

In the full tank state, adjust the gain so that the displayed value matches the actual level. It is recommended to repeat 2-3 times to improve accuracy.

Safety Precautions

  • Before operation, a radiation safety license must be obtained and the operation must be performed by licensed personnel.
  • It is strictly prohibited to disassemble the source container without authorization to avoid radiation leakage.
  • Regularly check the dose rate on the surface of the container to ensure it is below the limit.
  • Discarded source containers should be handed over to professional institutions for disposal.

Common Problems

  • Large signal fluctuation: Check grounding and shielding, or recalibrate.
  • No signal output: Confirm power supply and fuse, and check whether the source is in place.
  • Abnormal dose rate: Immediately isolate the area and contact the manufacturer.

For more details, please refer to the complete version of the VEGASOURCE 84 Chinese Manual.