Code

error description (display)

Problem Resolution

A101

A102

A110

A152

A160

Diagnostic Error

Reset

Alarm still present Replace circuit board

W103

is initialized

alarm information does not disappear within a few seconds to replace the circuit grid

A 106

when downloading

wait for the message to disappear after the download ends

A111

A113

A114

A115

A121

A125

A155

A164

A171

Circuit board failure

Reset

To monitor the system's electromagnetic compatibility (EMC) needs to be improved when the alarm remains after reset to replace the circuit board

A116

Download Error

Check connection

Download again

W153

is initialized

Wait a few seconds if the error persists Power off before powering up

A231

sensor fault

Check the connection If you need to replace the HF module or circuit board

A281

temperature probe broken

Road

Replace sensor

A502

Sensor Model

Not detected

Replace sensor and/or circuit board

W511

No factory calibration

for factory calibration

A512

do curve records

When

alarm disappears after a few seconds

A521

new transmission detected

Sensor model

Reset

W601

linear curve non

Monotonous

Correction data table (input monotonically increasing data table)

W611

less than two lines

sexual point

entering more data on

W621

Analog function ON

Close simulation function"output"(06) function group "mode

Proposed "(065) function

E641

No echo available

inspection (see page 21)

E651

liquid level in safety

Hazard of overflow within distance

liquid level disappears from the safe distance error may need to be reset.

Lock "Security Settings" (01) function group "Acknowledge Alarm" (017) function

A661

sensor temperature too high

Check the medium temperature or detection sensor

A671

Linearization Incomplete

whole

Activate Linearization Table

W681

Current out of range

Basic calibration check linearization

W691

W691

detected full noise Level ramp up activated