Instrumentation · Engineering · Package · PLC
0755-8222 1901 sales@ainstru.com

8572400

8572400

8572400

HACH 8572400
CL17sc 余(总)氯分析仪,带压力调节器安装工具包,无试剂 CL17sc Colorimetric Chlorine Analyzer with Pressure Regulator Installation Kit, w/o Reagents
Accuracy ± 5% or ± 0.04 mg/L (whichever is greater) from 0 - 5 mg/L Cl₂
± 10% from 5 - 10 mg/L Cl₂
Air Purge Optional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar (20 psig) maximum
Alerts Low and high chlorine. Dirty cell. Low and high sample flow.
Certifications CE compliant with: EN 61326-1, CISPR 11, EN 50581
ACMA RCM
South Korea KC Certificate
North America: FCC Supplier’s Declaration of Conformance, IEC/EN 60529, ICES-003
Controller Compatibility SC200, SC1000, SC4500. All controllers sold separately
Cycle Time 2.5 minutes
Dimensions (H x W x D) 342 mm x 329 mm x 177 mm
Drain ½-inch ID flexible hose
Enclosure Rating IP66
Inlet ¼-inch OD polyethylene tube, quick-disconnect fitting
Inlet Pressure 0.3 - 5.2 bar (4.5 - 75 psig) supplied to Y-strainer
0.1 - 0.3 bar (1.5 - 5 psig) supplied to analyzer
Installation Kit Pressure Regulator Installation Kit
Interferences Other oxidizing agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. Hardness must not exceed 1,000 mg/L CaCO₃.
Light Source LED, measurement at 510 nm, 1 cm light pathlength
Lower Limit of Detection (LOD) 0.03 mg/L
Mounting Wall mount
Operating Humidity 0 - 90% relative humidity, non-condensing
Operating Temperature Range 5 - 40 °C (41 - 104 °F)
Parameter Chlorine, free; Chlorine, total residual
Power Requirements (Voltage) 12 VDC, 400 mA maximum (supplied by the controller)
Range 0 -10 mg/L Cl₂
Reagent Consumption 0.5 L of buffer solution and 0.5 L of indicator solution in 31 days
Sample Flow Rate 60 - 200 mL/min through the instrument
Sample Requirements Y-strainer filtration with 40-mesh screen or higher
Sample Temperature 5 - 40 °C (41 - 104 °F)
Warranty 12 months
Weight 4.1 kg (9.0 lbs.)
What's included? CL17sc Chlorine Analyser with Pressure Regulator Installation Kit, and User Manual. User provides screws and/or anchors for mounting.
Hach SC Controller and Reagents are not included, they are sold separately.
Municipal Wastewater Treatment

Municipal Wastewater Treatment — Wastewater Treatment Plant · Hach Application Case

Project Background

Municipal Wastewater Treatment customers at the Wastewater Treatment Plant stage have long faced the challenge of insufficient accuracy in effluent quality monitoring and instruments prone to contamination causing data drift, affecting compliance discharge, placing higher demands on the stability and measurement accuracy of Hach.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Hach for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Municipal Wastewater Treatment conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Food & Beverage

Food & Beverage — Food Production Line · Hach Application Case

Project Background

Food & Beverage customers at the Food Production Line stage have long faced the challenge of high hygienic measurement requirements, with inline instruments needing to withstand CIP/SIP cleaning while maintaining stable accuracy, placing higher demands on the stability and measurement accuracy of Hach.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Hach for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Food & Beverage conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Chemical & Pharmaceutical

Chemical & Pharmaceutical — Reactors and Storage Tanks · Hach Application Case

Project Background

Chemical & Pharmaceutical customers at the Reactors and Storage Tanks stage have long faced the challenge of highly corrosive media and complex conditions, requiring corrosion-resistant and explosion-proof instruments, placing higher demands on the stability and measurement accuracy of Hach.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Hach for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Chemical & Pharmaceutical conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Power & Energy

Power & Energy — Power Plant Boiler and Turbine · Hach Application Case

Project Background

Power & Energy customers at the Power Plant Boiler and Turbine stage have long faced the challenge of unstable measurement under high temperature and pressure, with high maintenance and downtime costs, placing higher demands on the stability and measurement accuracy of Hach.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable Hach for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Power & Energy conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.