E-H Electromagnetic Flowmeter 10D Product Overview

The E-H electromagnetic flowmeter 10D is a flow measurement instrument based on Faraday's law of electromagnetic induction, suitable for volumetric flow measurement of conductive liquids. It features high measurement accuracy and good stability, and is widely used in water treatment, chemical, food and beverage industries. This article provides concise operating instructions and a selection guide for the 10D to help engineering technicians and procurement personnel get started quickly.

Download 10D Chinese Manual.pdf

10D Electromagnetic Flowmeter Technical Parameters

The following are the core technical parameters of the E-H electromagnetic flowmeter 10D for selection and installation reference.

ParameterSpecification
Nominal diameter rangeDN10 ~ DN2000
Accuracy class±0.5% or ±0.2% (optional)
Medium conductivity≥5 µS/cm
Working pressure≤4.0 MPa (depending on nominal diameter)
Medium temperature-20°C ~ +180°C (depending on lining)
Output signal4-20mA, pulse, HART, RS485
Protection classIP65, IP67, IP68 (optional)

10D Electromagnetic Flowmeter Installation Points

Correct installation is a prerequisite for ensuring measurement accuracy. Please follow the steps below for installation.

  1. Ensure the pipeline is filled with liquid to avoid air bubble accumulation.
  2. The sensor should be installed at the low end of the pipeline or on a rising pipe section to ensure full pipe.
  3. Ensure the length of straight pipe sections before and after: 5D before, 3D after (D is the pipe diameter).
  4. The grounding ring or grounding electrode must be reliably grounded, with a grounding resistance less than 10Ω.
  5. Avoid installation near strong electromagnetic interference sources, such as frequency converters and large motors.

10D Electromagnetic Flowmeter Parameter Setting and Commissioning

Parameter setting can be done through the converter panel or hand operator. Common settings include nominal diameter, range, output mode, etc.

Basic Setting Steps

  1. Enter the menu, select nominal diameter and range to ensure consistency with actual conditions.
  2. Set the output signal type (4-20mA or pulse).
  3. Enable empty pipe detection and small signal cut-off functions as needed.
  4. Save parameters and exit, observe whether the flow display is normal.

10D Electromagnetic Flowmeter Common Troubleshooting

The following lists common fault phenomena and handling methods for quick maintenance.

Fault phenomenonPossible causeHandling method
No flow displayPower not connected, fuse blownCheck power and fuse
Large flow display fluctuationAir bubbles, poor grounding, interferenceVent air, check grounding, stay away from interference sources
Large measurement deviationIncorrect nominal diameter or range setting, electrode contaminationReset parameters, clean electrodes
Abnormal output signalIncorrect wiring, excessive loadCheck wiring, reduce load resistance

10D Electromagnetic Flowmeter Selection and Maintenance Recommendations

When selecting, determine the lining and electrode materials based on medium characteristics, temperature, pressure, and nominal diameter. Regular maintenance can extend service life.

  • Lining selection: Rubber is suitable for water, PTFE is suitable for corrosive media.
  • Electrode selection: 316L is used for general water, Hastelloy is used for strong corrosion.
  • Maintenance cycle: It is recommended to check electrodes and grounding every 6 months.
  • Calibration: Perform online or offline calibration once a year.

For detailed information, please download the 10D Chinese Manual.pdf.