CA72TOC Analyzer Spare Parts Overview

CA72TOC is an analyzer from Endress+Hauser for online total organic carbon (TOC) monitoring, widely used in the pharmaceutical, food and beverage, and semiconductor industries. To ensure long-term stable operation of the instrument, regular replacement of spare parts is essential. This article compiles the commonly used spare part order numbers, specifications, and maintenance levels for the CA72TOC, and supplements the selection information for pH electrodes and reagents, making it convenient for engineering technicians and procurement personnel to quickly look up.

CA72TOC Analyzer Spare Parts Diagram

Instructions for Using the Spare Parts List

The "Maintenance Level" in the spare parts list is based on the SFT classification, indicating the skill level required to replace that spare part. Please select the appropriate spare parts according to your own qualifications to avoid instrument damage or safety risks caused by improper operation.

  • Category 1: Requires basic professional knowledge and can be replaced by yourself.
  • Category 2: Must complete E+H basic training before replacement can be performed.

CA72TOC Spare Parts Detail List

The table below lists the main spare parts of the CA72TOC analyzer, including order numbers, specifications, and maintenance levels. Some spare parts appear repeatedly, indicating that they are suitable for different positions or different versions; please refer to actual needs.

Order Number Specification Maintenance Level
71553782CA72TOC hose clamp
71556670Kit CA72TOC head with rolls for pump
71092635CA72xx fuse assembly
71450809Kit CA72TOC relay and fuses
71101602CA72xx pH electrode cable
71101610CA72xx diaphragm pump
71101611CA72xx diaphragm pump
71101615CA72xx tube
71101616CA72xx tube
71101532CA72TOC: hot salt collector assembly
71250117CA72TOC heated salt trap
71341850CA72TOC spare part: mixing chamberCategory 1
71101538CA72TOC flow meter gauge
71101548CA72TOC acid condensate adapterCategory 2
71101555CA72TOC acid adapterCategory 2
71101557CA72TOC sample adapterCategory 2
71101572CA72TOC tube
71101578CA72TOC: consumable tube
71101579CA72TOC: fuel inlet line II
71101580CA72TOC: fuel inlet line I
71101581Combustion furnace outlet (hot filter) assembly
71101582Combustion furnace outlet assembly
71101584CA72TOC syringe
71101585CA72TOC acid dispenser
71101586CA72TOC membrane filter
71101606CA72TOC acid filter separator
71101607CA72TOC acid filter
71101613CA72TOC tube
71101614CA72TOC tubeCategory 2
71101617CA72TOC connector
71101618CA72TOC O-rings and seals
71102294CA72TOC heater
71102295CA72TOC heater
71102317CA72TOC repair tools
71095149CA72TOC ammonium carbonate
71095156CA72TOC ammonium carbonate
71095158CA72TOC ammonium carbonate-free
71144072CA72TOC capillary
71105994CA72TOC FCS tube
71092159Repair part
71092160Repair part
71206103CA72xx maintenance
71232256Kit CO2 scrubber, annual consumable
71232261Kit CO2 scrubber, adsorber
71232262Kit CO2 scrubber, filter pad
71250117CA72TOC heated salt trap
71312862CA72TOC spare part: water supply pressure switchCategory 2
71295731CA72TOC spare part: water supply 24V
71304484CA72TOC spare part: water supply filter screen
71341850CA72TOC spare part: mixing chamberCategory 1
71363638CA72TOC spare part: MV5Category 2
71363643CA72TOC spare part: relay 2+8Category 2
71414588Kit CA72TOC Restrictor carrier gas
71414589Kit CA72TOC Restrictor stripp gas
71440164Kit CA72TOC Pump controller typ III
71440885Kit CA72TOC Gas supply typ III

pH Electrode and Reagent Selection

In addition to mechanical and electrical spare parts, the CA72TOC analyzer also requires regular replacement of pH electrodes and reagents to ensure measurement accuracy.

pH Electrode CPS71

CPS71 is a pH combination electrode suitable for hygienic processes, using a ceramic diaphragm and gel reference system, and can withstand high-temperature sterilization.

  • Diaphragm: ceramic
  • Reference system: gel (non-acrylamide), dual chamber
  • Application: hygienic, aseptic processes
  • Features: sterilizable, autoclavable
  • Minimum conductivity: 10 μS/cm

pH Standard Solution CPY20

CPY20 is a quality buffer solution traceable to PTB and NIST standard reference materials, supplied with a calibration certificate, suitable for calibration of pH electrodes.

Stripping Reagent CAY450

CAY450 is used for stripping inorganic carbon, contains hazardous substances, and must be handled in accordance with safety regulations.

Standard Solution CAY451

CAY451 is used to prepare standard solutions, also contains hazardous substances, please handle with care.

CA72TOC pH Electrode CPS71

Maintenance Precautions

When replacing spare parts, be sure to observe the following principles to ensure personnel safety and instrument performance.

  1. Confirm the maintenance level of the spare part; Category 2 spare parts must be replaced by personnel who have completed E+H basic training.
  2. Before replacement, turn off the power and gas supply, and release the pressure in the lines.
  3. Use original spare parts to avoid measurement deviation caused by compatibility issues.
  4. After replacement, perform calibration or verification to ensure the analyzer operates normally.

Frequently Asked Questions

How to determine the maintenance level of a spare part?

The spare parts list indicates Category 1 or Category 2. If not indicated, it is usually a regular spare part and can be replaced by yourself.

Can the CO2 scrubber kit be used for Parker adsorbers?

No. The Kit CO2 scrubber series (order numbers 71232256, 71232261, 71232262) is explicitly not suitable for Parker's CO2 adsorbers.

What accessories are needed for the pH electrode CPS71?

It must be used with the pH electrode cable (order number 71101602) and calibrated regularly with CPY20 standard solution.