Prosonic 93 Cumulative Flow Reset and Clearing Operation Guide

Resetting and clearing the cumulative flow of the Prosonic 93 is a common operation in daily maintenance. Correctly understanding its display mechanism and operation steps can avoid data misreading and operational errors. This article is intended for engineering and technical personnel and procurement personnel, providing clear operational guidance.

Prosonic 93 cumulative flow display interface

Cumulative Flow Display Mechanism

The cumulative flow value of the Prosonic 93 consists of up to 7-digit floating-point numbers. When the cumulative amount exceeds 9,999,999, the instrument extends the display range through the OVERFLOW function. The effective total value is the cumulative amount displayed by the OVERFLOW function plus the value of the SUM function.

Overflow Value Calculation Example

The following example illustrates how to combine the OVERFLOW and SUM functions to read the true cumulative flow:

FunctionValueDescription
OVERFLOW2E7 dm³i.e., 20,000,000 dm³
SUM196,845 dm³Current cumulative value
Effective total value20,196,845 dm³OVERFLOW + SUM
Prosonic 93 overflow and cumulative function display

Cumulative Flow Reset and Clearing Steps

Before performing reset and clearing, please confirm that the current cumulative value has been recorded to avoid losing important data. The operation steps are as follows:

  1. Enter the instrument main menu and select the "Cumulative Flow" or "Totalizer" option.
  2. Find the "Reset" function and press the confirm key to enter.
  3. The system prompts for reset confirmation; select "Yes" or "OK" to complete clearing.
  4. Check whether the SUM function display returns to zero and whether the OVERFLOW value is reset simultaneously.

Precautions

  • The reset operation may affect historical data records; it is recommended to perform it during batch switching or maintenance.
  • If the instrument is password-protected, the correct password must be entered first.
  • After reset, the cumulative flow starts recalculating from zero.

Mastering the Prosonic 93 cumulative flow reset and clearing method can ensure accurate and reliable flow measurement. For more technical support, please refer to the product manual or contact the supplier.