8BE Canning Mass Flowmeter Spare Parts List
The 8BE canning mass flowmeter is widely used in filling applications. The correct selection of its spare parts and accessories is crucial for maintenance and debugging. The following spare parts table lists commonly used spare parts and order numbers for easy procurement and inventory management.
| No. | Order Number | Specification Description |
|---|---|---|
| 3 | 50106219 | Fuse |
| 2 | 50104802 | Housing cover seal |
| 8 | 50106443 | FXA service adapter Batchline packed |
| 8 | 71035809 | 5BH service cable Y-cable |
| 7 | 50107167 | Adapter |
| 6 | 50107168 | Adapter |
| 5 | 50107169 | Adapter |
| 4 | 50107895 | Connecting cable RKWTN8-56/5 P92, 5m |
Communication Interface FXA291
FXA291 is a communication interface used to connect a computer USB port to the instrument service interface. It supports USB 1.1 or higher and is powered via USB. Its protection rating is IP51; NEMA/Typ 3, ambient temperature range -20...+60°C, and it has Ex: II(2) G (EEx ia) II C and ATEX: BVS 05 ATEX E 095 certifications.
Scope of Delivery
- Commubox FXA 291
- MS Windows USB driver
This device has been submitted for CSA/FM application certification. Note: The device must not be modified and may only be used in intrinsically safe circuits with corresponding characteristics (see safety certificate).
8BE Mass Flowmeter Selection Code Description
The 8BE mass flowmeter is a canning mass flowmeter OEM version, with a stainless steel housing, suitable for filling applications (compact type). The following is the complete selection code and option description.
Nominal Diameter
- 08: DN08 3/8"
- 15: DN15 1/2"
- 25: DN25 1"
Measuring Tube Material
- 9: Special version, TSP number required
- A: Stainless steel
- C: Stainless steel, with material certificate (wetted parts)
Process Connection
| Code | Description |
|---|---|
| 999 | Special version, TSP number required |
| D2S | PN40, 316L/1.4404, EN1092-1 B1 DIN2501 flange |
| FDD | DIN32676, 0.4um 3A, 316L/1.4404 |
| FJA | ISO2853, 0.8um/150grit 3A, 316L/1.4404 |
| FJW | ISO2853, 316L/1.4404 |
| FLA | DIN11864-1A, 0.8um/150grit 3A, 316L 1.4404 |
| FLW | DIN11864-1A, 316L/1.4404 |
| FMA | DIN11851, 0.8um/150grit 3A, 316L 1.4404 |
| FMW | DIN11851, 316L/1.4404 |
| FSA | SMS1145, 0.8um/150grit 3A, 316L/1.4404 |
| FSW | SMS1145, 316L/1.4404 |
| FTA | Tri-Clamp, 0.8um/150grit 3A 316L/1.4404 |
| FTD | Tri-Clamp, 0.4um 3A, 316L/1.4404 |
| FTS | Tri-Clamp, 316L/1.4404 |
| FUA | Tri-Clamp 1/2", 0.8um/150grit 3A, 316L 1.4404 |
| FUD | Tri-Clamp 1/2", 0.4um 3A, 316L/1.4404 |
| FUW | Tri-Clamp 1/2", 316L/1.4404 |
| FWA | Tri-Clamp 3/4", 0.8um/150grit 3A, 316L 1.4404 |
| FWW | Tri-Clamp 3/4", 1.4404/316L |
Additional Testing
- 9: Special version, TSP number required
- A: Basic version
- B: Pressure test (piping)
- C: Degreasing and de-oiling cleaning (wetted parts)
- D: Degreasing and de-oiling cleaning (wetted parts) + pressure test (piping)
Calibration
- 0: No calibration (±5% o.r.)
- 9: Special version, TSP number required
- A: 0.3% + confirmation
- B: 0.15% + protocol
Certification
- 9: Special version, TSP number required
- A: Non-hazardous area
- C: cCSAus Class I, Div.2
- S: ATEX II 3G Ex nA IIC
Housing
- 9: Special version, TSP number required
- B: Compact IP67 NEMA4X, stainless steel
Cable Entry
- 4: Direct connection RSE8, without adapter 24VDC, pulse, status, service
- 9: Special version, TSP number required
Power Supply
- 1: 20-30 VDC
- 9: Special version, TSP number required
Software Function
- 9: Special version, TSP number required
- A: Basic version
Output/Input
- 3: 2x pulse/frequency/status output, max 10kHz; passive
- 4: Modbus, 1x batch output, status input
- 5: Modbus, 2x batch output, status input
- 9: Special version, TSP number required
Marking
- Z1: Tag, see additional specifications
Communication Interface FXA193
FXA193 service interface is used to connect the service port of Proline and ToF devices to the RS 232C serial interface (9-pin) of a personal computer.
Scope of Delivery
- Service interface with RS 232C connecting cable (9-pin)
- FXA "USB" cable for providing additional power to FXA193 from the PC's USB port
- Service software ToF Tool-Fieldtool package
Certification Options
- 9: Special version, TSP number required
- A: For non-hazardous area
- B: ATEX II (1) GD
- C: CSA/FM Class I Div. 1
- D: ATEX, CSA, FM
Connecting Cable Options
- 9: Special version, TSP number required
- A: Proline devices
- B: ToF devices
- E: Proline and ToF devices
- F: Ex two-wire devices
- G: Proline devices and Ex two-wire devices
- H: Proline, ToF and Ex two-wire devices
- X: Without connecting cable