Promag 400 Alarm Code Overview
Promag 400 is a high-performance electromagnetic flowmeter launched by Endress+Hauser and is widely used in industrial process measurement. Its alarm code system helps users quickly identify fault types and guides maintenance. This article summarizes common Promag 400 alarm codes and their handling methods.

Common Alarm Codes and Handling
Sensor-Related Alarms
- S001: Sensor signal exceeds limit; check whether the electrodes are contaminated or the pipe is empty.
- S002: Excitation current abnormal; check coil connections and insulation.
- S003: Sensor temperature too high; check the medium temperature and ambient temperature.

Transmitter-Related Alarms
- C001: Internal communication fault; restart the transmitter or contact the manufacturer.
- C002: Power supply voltage abnormal; check whether the supply voltage is within range.
- C003: Memory error; the electronic module needs to be replaced.

Empty Pipe Detection Alarms
- E001: Empty pipe; check whether the medium fills the pipe.
- E002: Empty pipe detection threshold set improperly; recalibrate.

Alarm Code Troubleshooting Steps
- Record the alarm code and the operating conditions at the time of occurrence.
- Refer to the manual to confirm the meaning of the alarm.
- Carry out preliminary troubleshooting according to the handling recommendations provided in this article.
- If the problem cannot be solved, contact Endress+Hauser technical support.

Alarm Code Quick Reference Table
| Alarm Code | Description | Possible Cause | Handling Measure |
|---|---|---|---|
| S001 | Sensor signal exceeds limit | Electrode contamination, empty pipe | Clean electrodes, check medium |
| S002 | Excitation current abnormal | Coil fault | Check coil connections |
| C001 | Internal communication fault | Electronic module fault | Restart or replace module |
| E001 | Empty pipe alarm | Pipe not full | Adjust process |

Preventive Maintenance Recommendations
- Regularly inspect electrodes and liners to prevent scaling and corrosion.
- Ensure good grounding to avoid electromagnetic interference.
- Calibrate regularly to ensure measurement accuracy.

Summary
Mastering the meanings and handling methods of Promag 400 alarm codes can effectively reduce downtime and improve production efficiency. For more detailed fault diagnosis, please refer to the official manual or contact technical support.